Objective:
To develop a smart contact lens for non-invasive ocular blood oxygen saturation monitoring.
Approach:
- Device Design: A flexible plasmonic nano-confinement nanowire photodetector integrated into a soft contact lens, adapting pulse oximetry principles for ocular use.
- Measurement Technique: Utilizes red and near-infrared LEDs to illuminate the eyelid and captures photoplethysmography signals from eyelid capillaries.
- Nanowire Architecture: Employs a PEDOT:PSS nanowire array embedded with gold nanoparticles to enhance organic photodetector performance.
- Testing and Validation: Initial testing showed high agreement with commercial oximeters, and evaluations in anesthetized rabbits demonstrated parallel SpO₂ readings.
Key Findings:
- The smart lens achieved approximately 99 percent average agreement with a commercial finger-clip oximeter.
- The PNC nanowire architecture improved test accuracy by 35 percent compared to sensors without it.
- Eyelid SpO₂ readings changed in parallel with peripheral SpO₂ under varying oxygen conditions.
Interpretation:
The device maintained high transmittance and mechanical stability, showing promise for continuous monitoring of ocular oxygen saturation.
Limitations:
- Validation covered an SpO₂ range of roughly 84–100 percent, excluding severe hypoxaemia.
- Testing was conducted in anesthetized animals, minimizing blinking and eye-motion artefacts.
Conclusion:
Future work will focus on improving lens stability, long-term exposure to tear fluid, and integrating wireless operation.
Sources:
This content is an AI-generated, fully rewritten summary based on a published scholarly article. It does not reproduce the original text and is not a substitute for the original publication. Readers are encouraged to consult the source for full context, data, and methodology.